top of page

The Clean Air'Ambient Laser Dust Monitor. Measurement flow rate : 0.1cfm 2.83LPM Measurement size: pm10, pm2.5, pm1 Digital Output: TCP/IP, RS-232 Particle separation : Pm10 impactor & Optical isolation Measure Pm10, pm2.5, pm1.0 at the same time

 

Dust Monitoring System | 微塵分析 PM1, PM2.5

  • MODEL : LDU-T100 

                       LDU-100 

                       LDU-125 

                       LDU-125M

    Measurement method : Light scattering

    Light source : Laser Diode classⅢ

    Particulate matter : TSP 

                        PM10 

                        PM2.5, PM10 PM1.0, 

                        PM2.5, PM10

    Measurement range : 0~10000㎍/㎥ 

                        PM 1.0:0~500㎍/㎥, PM2.5:0~750㎍/㎥, PM10:0~2000㎍/㎥

    Resolution : ±2㎍/㎥ 

                       ±0.3㎍/㎥

    Sensitivity : 1㎍/㎥ 

                       0.1㎍/㎥

    Particle size range : 0.3㎛~15㎛

    Flow rate : 0.5 LPM

    Data upload : 10 sec (1 min cumulation data)

    Display : Graphic LCD 128 x 64

    Network interface : IEEE 802.3 Ethernet TCP/IP v4

                                            Option: IEEE 802.11b/g Wireless LAN,

                                                             IEEE 802.15.1 Bluetooth v2.0 Class2(ISM Band, 2.4 to 2.485GHz)

    Operating environment : 14℉~122℉ (-10℃~50℃), 20~75% RH

    Storage environment : -4℉~140℉ (-20℃~60℃), ≤98% RH

    Weight : Body 844g / Body & Support 972g

    Dimension(W x H x D) : 160㎜ x 175(230)㎜ x 65(139)㎜ (with inlet and bottom support)

    Power : 12VDC, 450mA

    Battery(optional) : Li-polimer (max. 4 hour)

    Calibration : Labco calibration system in accordance with ISO-21501-4

    Calibration interval : One year or less (ISO 21501-4)

bottom of page