The Clean Air'Ambient Laser Dust Monitor. Measurement flow rate : 0.1cfm 2.83LPM Measurement size: pm10, pm2.5, pm1 Digital Output: TCP/IP, RS-232 Particle separation : Pm10 impactor & Optical isolation Measure Pm10, pm2.5, pm1.0 at the same time
Dust Monitoring System | 微塵分析 PM1, PM2.5
MODEL : LDU-T100
LDU-100
LDU-125
LDU-125M
Measurement method : Light scattering
Light source : Laser Diode classⅢ
Particulate matter : TSP
PM10
PM2.5, PM10 PM1.0,
PM2.5, PM10
Measurement range : 0~10000㎍/㎥
PM 1.0:0~500㎍/㎥, PM2.5:0~750㎍/㎥, PM10:0~2000㎍/㎥
Resolution : ±2㎍/㎥
±0.3㎍/㎥
Sensitivity : 1㎍/㎥
0.1㎍/㎥
Particle size range : 0.3㎛~15㎛
Flow rate : 0.5 LPM
Data upload : 10 sec (1 min cumulation data)
Display : Graphic LCD 128 x 64
Network interface : IEEE 802.3 Ethernet TCP/IP v4
Option: IEEE 802.11b/g Wireless LAN,
IEEE 802.15.1 Bluetooth v2.0 Class2(ISM Band, 2.4 to 2.485GHz)
Operating environment : 14℉~122℉ (-10℃~50℃), 20~75% RH
Storage environment : -4℉~140℉ (-20℃~60℃), ≤98% RH
Weight : Body 844g / Body & Support 972g
Dimension(W x H x D) : 160㎜ x 175(230)㎜ x 65(139)㎜ (with inlet and bottom support)
Power : 12VDC, 450mA
Battery(optional) : Li-polimer (max. 4 hour)
Calibration : Labco calibration system in accordance with ISO-21501-4
Calibration interval : One year or less (ISO 21501-4)