top of page

The Class 1 IoT data logger meets all requirements for noise measurement and frequency analysis. The IoT meter is equipped with a GPS receiver in addition to the sound measurement function. The GPS function allows the user to assign the measured values to an exact position.

This integrated precision pulse SPL meter follows all valid standards and guidelines (Class 1 EN/IEC 61672, ANSI S1.4-1983, ANSI S1.43-1997 EN/IEC61260, etc.) The IoT meter has a very large illuminated display. The SPL meter display adjusts the sound pressure level numerically and graphically in real time. The professional SPL meter has an integrated data logger. The measurement data is stored on a micro SD card. This card can be read directly in the PC. 

Likewise, the IoT data logger can be connected to the PC via USB and the memory can be read out. The post-processing software from the sound level meter allows convenient measurement data analysis. The typical applications for this instrument are the measurement of workplace noise, compliance, environmental noise detection, selection of hearing protector selection, selection of appropriate noise abatement measures and noise exposure estimation.

The pulse function of the IoT data logger allows you to record short-term sound events. Thanks to the integrated octave band function even the smallest differences in the frequency band can be detected. The frequency filter can be optionally upgraded. The measuring range of this IoT meter ranges from 22 ... 136 db (A) at a frequency of 3 Hz ... 20 kHz.

PCE-4xx-EKIT
An additional kit for IoT data logger can be combined with the SPL meters PCE-428, 430 or PCE-432. This kit makes it possible to measure the outdoor noise with a noise meter over a long period of time. The environmental noise monitoring system consists of a waterproof Peli carrying case with rollers. A charger and two additional lead gel batteries are installed in this case, which allows an operation of the sound monitor for up to 10 days. During the IoT meter, the case may remain closed. Due to the external connections for the microphone and the power supply all electronic components are protected from water.

The delivery scope of the IoT data logger includes rain and wind protection as well as a wire mesh for the microphone. Also, a tripod is available with which the microphone can be placed outdoors. Also, the charger in the measuring kit PCE-4xx-EKIT is waterproof. So, the charger may also remain outdoors. The IoT meter can be used, for example, for the sound measurement of road noise, aircraft and train noise, event noise, etc. The weather does not matter, hail, snow and rain are no problem for the kit for outdoor noise. If the IoT meter is combined with this measuring kit, it is even possible that the exact location of the sound level metering around the outside area is assigned via the GPS receiver.

 

Datasheet

IoT Data Logger PCE-432-EKIT

數量
  • 1/1 octave band included

    - 1/3 octave band optional

    - Accuracy class 1 

    Integrated GPS receiver 

    - A, B, C & Z frequency rating

    - Fast, Slow, Pulse and Peak time weighting

    - Statistics function

    - Display of the sound profile as a graph

    - 3 measuring profiles adjustable

    - Low inherent noise

    - ICCP microphone 40 mV /PA

    - Alarm level adjustable
     

  • Measuring range

    22 ... 136 db(A)

    Accuracy

    Class 1

    Frequency range

    3 Hz ... 20 kHz

    Standards

    GB/T 3785.1-2010
    GB/T 3785.2-2010
    IEC 60651:1979
    IEC 60804:2000
    IEC 61672-1:2013
    ANSI S1.4-1983
    ANSI S1.43-1997

    Frequency analysis

    1/1 Octave band filter: 8 Hz ... 16 kHz
    1/3 Octave band filter: 6.3 Hz ... 20 kHz

    Microphone

    1/2"
    Microphone Class: 1
    Sensitivity: 40 mV/PA
    Frequency range: 3 Hz ... 20 kHz
    Connection: TNC
    Power supply: ICCP Standard

    Data-logging interval

    1 s ... 24 h (adjustable)

    Measuring functions

    LXY (SPL), LXeq, LXYSD, LXSEL, LXE, LXYmax, LXYmin, LXPeak, LXN

    X = Frequency weighting: A, B, C, Z
    Y = Time weighting: F, S, I
    N = Statistics in %: 1 ... 99 %

    Frequency weightings

    A, B, C, Z

    Time weightings

    Fast (F) 125ms, Slow (S) 1 sec, Impulse (I) 35 ms

    Inherent noise

    Microphone: 19 db(A), 25 db(C), 31 db(Z)
    Electronics: 13 db(A), 17 db(C), 24 db(Z)

    AD converter

    24 Bit

    Sample rate

    Standard: 48 kHz
    LN Mode: 20 ms

    Measuring display

    Numerical
    Bar graph
    Graphical

    Display

    160 x 160 pixel LCD with backlight

    Memory

    4 GB Micro SD card

    Interface

    USB (Memory readable via software or directly as mass storage)
    RS-232

    Voltage output

    AC 5V RMS
    DC 10 mV/db

    Alarm

    Adjustable

    GPS Yes, GPS receiver for location determination

    Power supply

    4 x 1.5 V AA Batteries
    12 V / 1 A Power plug
    5 V / 1 A USB

    Battery life

    Min. 10 h continuous use

    Dimensions

    70 x 300 x 36 mm / 2.76 x 11.81 x 1.42" (W x H x D)

    Weight

    Approx. 620 g / 1.4 lbs incl. batteries

Copyright© 2024 by Tops Instruments Supplies Co.  graphics, data sheet, photos and videos all rights reserved. 