Our Moisture Analyser / Moisture Analyzer has been specially developed for light and bulky materials. It has an exceptionally large drying pan and is suitable for applications in the fields of raw material extraction, production, processing, research and recycling.
With a maximum weighing capacity of 400 g, a volume of 360 cm³ and a high measuring resolution of 0.001 g, the Moisture Analyser / Moisture Analyzer enables precise and reliable moisture measurements. The balance works according to the principle of drying with simultaneous weighing: the weight loss is continuously recorded and displayed as a percentage of the total mass. With a heater output of 2 x 400 W and an adjustable heating temperature of 105 °C to 360 °C, the Moisture Analyser / Moisture Analyzer offers quick and flexible drying of samples. The device provides various parameters for analysis, including % moisture, % dry matter, % moisture bd as well as g residual weight/kg and residual weight.
The practical functions include an automatic power-off function that stops the drying process when a constant weight is reached. This can be flexibly set to three criteria: sampling start, sampling interval and weight loss per sampling interval. A timer with an adjustable duration of 1 to 180 minutes is also available. A menu lock protects the parameters against unintentional adjustment; the menu can be locked by a PIN whereas the CAL menu item remains accessible. The balance offers up to 9 programme memory locations for easy handling of drying profiles.
Data is output via an RS-232 interface or a USB connection. An optional software is available for transferring data to a PC. GLP-compliant documentation is carried out via the Data Bridge as the balance does not have an integrated printer. However, an external printer can be purchased separately to output the measurement data directly. A 4-digit device ID also enables the unique allocation of the measurement data. The Moisture Analyser / Moisture Analyzer is ideal for applications in the wood, agricultural and plastics industries, e.g. for shavings, animal feed or granulates. It delivers precise results even at very low moisture contents.
Moisture Analyzer / Moisture Analyzer PCE-UX 3081WQ
- weighing range: 400 g
- sample weight: min. 20 g
- sample volume: max. 360 cm³
- weighing pan inside: 230 x 97 mm / 243 x 111 mm
- adjustable heating temperature: 105 ... 360 °C
- radiator power: 2 x 400 W
- time setting: 1 ... 180 min
- menu lock
Absolute moisture Measurement range up to 0 ... 100 % Resolution 0,01 % Accuracy 0.01 % Weighing range Measurement range up to [Max] 400 g Resolution [d] 0,001 g Accuracy 0.001 g General technical data Outputs V24 Measuring functions % moisture, % dry matter, % moisture bd, g residual weight / kg, Residual weight Display type LCD with backlight Display size 4 Inch Interface USB, RS232 Heating temperature 105 ... 360 °C Minimum load [min] 20 g Menu language German, English, English (GB) Protection class (device) IP20 Power supply 230V AC | 48/63 Hz Connector type Schuko plug Operating conditions 10 ... 40 °C , 0 ... 90 % RH Storage conditions 10 ... 40 °C , 0 ... 90 % RH Dimensions ( L x W x H x D ) 440 x 440 x 240 x 330 mm Other dimensions Inner pan size: 230 x 97 mm / 243 x 111 mm
Inner pan height: 17 mmWeight 13200 g