top of page

The moisture meter has a measuring range of 0 ... 100% absolute humidity. The moisture meter is operated via the 5" large touchscreen. To measure the moisture content with the moisture meter, a sample is evenly distributed on the sample pan. The sample pan is then placed and sealed in the moisture meter. After setting all parameters, the moisture meter begins the measurement.

 

The moisture meter determines the moisture content based on the measured weight difference, relative to the weight before and after drying the sample. The adjustable temperature range for the moisture meter is between 40 ... 165 °C. Thus, a specific drying temperature for the sample can be selected with the moisture meter. The generated drying temperature is produced by a built-in halogen lamp with a diameter of 90 mm.

 

The drying process of the sample makes the moisture meter usable in various fields. Thus, the moisture meter finds its application, for example, in laboratories to measure the moisture of different powders, in quality control to determine the moisture of granules, or on farms to determine the moisture content of oats, grains, or wheat.

 

A major advantage over devices using capacitive measuring methods is that the moisture meter operates independently of characteristic curves and provides more accurate results.

 

The measured values can be transferred to a PC via the Serial RS232 interface of the moisture meter for analysis. A suitable software is available optionally for the moisture meter. If required, the moisture meter can be supplied with an ISO calibration certificate.

 

Moisture Meter PCE-MA 200TS

  • - Measurement range from 0 ... 100%

    - Weighing range up to 200 g

    - High resolution of 0.01 g

    - Heating temperature from 40 ... 165 °C

    - Memory for 32 measurements

    - Suitable for various materials

  • Absolute Moisture  
    Measurement Range 0% ... 100%
    Resolution 0.01%
    Accuracy ± 0.04%
       
    Weighing Range  
    Measurement Range 0 g ... 200 g
    Resolution 0.01 g
    Accuracy ± 0.04 g
       
    General Technical Data  
    Measurement Functions ATRO Moisture Content %R
      Dry Content %D
      Moisture Content %M
      Initial Weight %W
    Display Type LC Touch Display
    Display Size 5 inches
    Display Refresh Rate 10 times per second
    Memory Capacity 32 data sets
    Data Interface RS232
    Heating Element Halogen Lamp
    Heating Temperature 40 ... 165 °C
    Measurement Rate 1 Hz
    Abort Criteria Weight Loss, Time
    Menu Language Chinese, Russian, Polish, Turkish, Portuguese, Dutch,
    Spanish, Italian, French, English, German
    Protection Class IP20
    Power Supply Schuko Plug (110 V - 240 V AC | 50/60 Hz)
    Weight 3.7 kg
    Device Weight with Package 4.3 kg
    Device Weight with Package and Outer Packaging   5.34 kg
    Dimensions (L x W x H) 375 x 225 x 205 mm
    Additional Dimensions Sample Pan Ø 90 mm, Height 6 mm
    Operating Conditions 20 ... 25 °C, 40 ... 60 % r.F.
    Storage Conditions 0 ... 40 °C, 0 ... 80 % r.F.

Copyright© 2024 by Tops Instruments Supplies Co.  graphics, data sheet, photos and videos all rights reserved. 